Electrical Requirements
The electrical rating for the 230 VAC unit is 230 VAC, 50/60 Hz, 10 amps. The
electrical rating for the 120 VAC unit is 120 VAC, 60 Hz, 15 amps. The electrical
rating for the 100 VAC unit is 100 VAC, 50/60 Hz, 20 amps. The three-pronged
grounding plug on the sterilizer power cord must be plugged into a matching
three-pronged, grounded, isolated, correctly polarized receptacle. Check the
serial number label on the back panel of the sterilizer to verify the voltage rating
for the unit.
DESCRIPTION
Operational Features
The M11 UltraClave™ sterilizers . . .
will begin a sterilization program cycle when a sterilization program is
selected, the door is closed and fully latched, and the START switch is
pressed.
will automatically fill the chamber with water to the correct level once the
START switch is pressed.
begin the cycle time countdown only after the proper pressure and tempera-
ture have been reached.
automatically exhaust the steam pressure from the chamber and open the
door slightly to facilitate drying upon completion of the sterilization phase.
automatically starts a 30 minute drying phase. Items that do not require dry-
ing may be removed any time after the sterilization phase is complete.
Ste-
rility of unwrapped items and moist wrappings is compromised on
exposure to a non-sterile environment.
will signal the operator that the drying phase and therefore the sterilization
WARNING
Use 220 - 240 VAC, 50/60 HZ alternating current only for 230
VAC models and 90 - 110 VAC, 50/60 HZ alternating current
only for 100 VAC models. Failure to do so could result in electrical
shock to personnel and will result in damage to sterilizer.
Do not use this sterilizer in an explosive or oxygen-enriched atmo-
sphere. Failure to do so could result in serious personal injury or
death.
NOTE
Grounding reliability can only be achieved if this unit is connected to a match-
ing three-pronged, grounded, isolated, correctly polarized receptacle.