Error Code Guide
Use the following table to determine the meaning of an error code and assist in
correcting minor problems with the sterilizer.
sterilizer is not level level the sterilizer.
input voltage is too low. have a qualified electrician
connect sterilizer to a
separate (dedicated) circuit
with proper voltage level.
Error Code Guide
Problem Possible Cause Solution
Error Code 1
(power interruption to
sterilizer)
unit powered up for first
time.
press ON / STANDBY
button and then continue
normal operation.
power was interrupted by
power outage.
press ON / STANDBY
button and then continue
normal operation.
overheat switch was
activated.
allow sterilizer to cool
completely. Restart
sterilizer. If unit has errors
again, unplug unit and
contact an authorized
service representative (see
Calling For Service later in
this manual).
Error Code 2
(STOP button pushed
during cycle)
_ restart sterilizer after unit
stops beeping.
Error Code 3
(ON / STANDBY button
pushed during cycle)
_ restart sterilizer after unit
stops beeping.
Error Code 4
(door ajar during cycle)
Door was opened during
cycle.
restart sterilizer after unit
stops beeping.
door switch not functioning
properly.
allow sterilizer to cool
completely. Restart
sterilizer. If unit has errors
again, unplug unit and
contact an authorized
service representative (see
Calling For Service later in
this manual).