Ref. Control Function
1 Temperature / Time dis-
play
alternately indicates cycle temperature and
exposure time when a cycle is selected. Dur-
ing a cycle, the cycle time remaining is dis-
played.
2 Pressure display indicates current chamber pressure during a
cycle.
3 FILLING indicator illuminates when the chamber is being filled
with water at the beginning of a cycle.
4 STERILIZING indicator illuminates during the heat up and sterilization
phases of a cycle.
5 DRYING indicator illuminates during the drying phase of a cycle.
6 COMPLETE indicator illuminates when a cycle is complete, indicat-
ing that the sterilized goods may be removed
at this point.
7 WATER LOW indicator indicator flashes (and a beep sounds) when
the water level is too low in the reservoir to ini-
tiate a cycle.
8 DOOR AJAR indicator indicator flashes (and a beep sounds) when
the door is not completely latched.
9 ERROR indicator indicator flashes (and a beep sounds) when a
cycle is terminated before completion.
NOTE: a corresponding error code is dis-
played on the Temperature / Time display any
time an error occurs. See Error Code Guide
for a detailed explanation of each error code.
10 UNWRAPPED button when pressed, selects a program cycle
designed to process unwrapped instruments
at 132°C (270°F) for 3 minutes with a 30
minute drying cycle following the sterilization
phase.
11 POUCHES button when pressed, selects a program cycle
designed to process instruments and dental
handpieces in combination paper / plastic
sterilization pouches or wrapped instruments
at 132°C (270°F) for 15 minutes with a 30
minute drying cycle following the sterilization
phase.
12 LIQUIDS button when pressed, selects a program cycle
designed to process non-volatile liquids at
121°C (250°F) for 30 minutes.
13 PACKS button when pressed, selects a program cycle
designed to process packs of instruments at
121°C (250°F) for 30 minutes with a 30
minute drying cycle following the sterilization
phase.