AT-A-GLANCE
CALIBRATION PROCEDURE
1) Open door.
2) Depress LIQUIDS
and PACKS buttons at the same time as plugging in the
power cord.
3.) Starting with UNWRAP
button, depress and release all buttons from left
to right across panel.
4) Turn
pressure zero pot #1 counter-clockwise to 0.0 PSI. Then turn clock-
wise until the display pressure reads
0.1 PSI.
5) Depending on if the board is old or new, connect multi-meter leads as
shown in Figure 2 on page 13 (old), or Figure 3 on page 14 (new).
6) Adjust
temperature pot #3 to 2.56 VDC reading on multi-meter.
7) Connect the test pressure gauge, close the door and press the ST
ART
button.
8) Check for pressure leaks at this time. Refer to 1.0 Leak Detection Guide.
9) After display reaches 272˚F, adjust the
pressure range pot #2 until the
display pressure matches test pressure gauge to within +/- 0.5 PSI
10) Depress the STOP
button to end the test cycle.
11) After the sterilizer vents and the door opens, unplug the power cord
(main PC board to reset itself).
The pressure zero, pressure range, and temperature potentiometers must
not be adjusted exclusively. If one potentiometer needs to be adjusted, then
the entire potentiometer adjustment procedure must be performed, and it
must be performed in the proper sequence as described in the steps below.
INSTRUCTIONS
1) Remove the power cord from the A.C. outlet. Remove right hand
side panel.
2) Open the sterilizer door.
3) To begin diagnostics, go to page 10, see Figure 1.