21-02-0285 REV. 06 07/11 CARDELL® Veterinary Vital Signs Monitors Models 9401 and 9402 User’s Manual Manufactured for: Go To Table
21-02-0285 REV. 06 07/11 This page is intentionally left blank NextBackGo To Table Of Contents
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 100 OTHER ACCESSORIES: Catalog No. Description 01-02-0395 Replacement Power Cord, U.S.A.
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 101 MONITOR CONFIGURATIONS Table 8: Monitor Configurations Model Description 9401 MAXNIBP
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 102 This page is intentionally left blank NextBackGo To Table Of Content
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 103 12. GLOSSARY Diastolic The Diastolic pressure is the bottom (lower) number given in a
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 104 This page is intentionally left blank NextBackGo To Table Of Contents
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 105 13. SPECIFICATIONS NIBP MEASUREMENT Characteristic Specification Technique:Oscillom
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 106 PATIENT ALARMS 9401 9402 Patient Parameters Low Limit High Limit Low Limit Hi
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 107 POWER Characteristic Specification Source: External line or internal battery AC Power
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 108 OPERATING ENVIRONMENT Characteristic Specification Operating Temperature: 0°C to 50°C (
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 109 STANDARDS Units comply with the following requirements: IEC 60601-1 EN 60601-1-2 EN 6060
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 11 TABLE OF CONTENTS 1. INTRODUCTION AND INTENDED USE ...
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 110 This page is intentionally left blank NextBackGo To Table Of Contents
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 111 14. PURCHASING RECORD CARDELL VETERINARY MONITOR Model Purchased (select one): 9401
B-26© Midmark Corporation 2009Midmark Corporation60 Vista DriveVersailles, OH 45380-0286Phone: 937-526-3662Fax: 937-526-5542midmark.comBackGo To Tab
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 12 REAR PANEL ...
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 13 MONITOR CONFIGURATION ...
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 14 PNEUMATIC TUBING ...
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 15 FIGURES Figure 1: Patient Environment ...
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 16 TABLES Table 1: Parts of the System ...
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 17 1. INTRODUCTION AND INTENDED USE INTRODUCTION The CARDELL Veterinary Monitors, Models 9
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 18 On each detected pulse, the perfusion LED does indicate patient perfusion signals. This b
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 19 DEFINITION OF TERMS In this manual, “WARNING”, “CAUTION”, “IMPORTANT” and “NOTE” mean the
21-02-0285 REV. 06 07/11 NextBackGo To Table Of Contents
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 20 This page is intentionally left blank NextBackGo To Table Of Contents
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 21 2. UNPACKING THE MONITOR INITIAL INSPECTION Before unpacking the monitor, inspect the
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 22 This page is intentionally left blank NextBackGo To Table Of Contents
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 23 3. SYMBOLS Units may display the following symbols: Alternating Current CAUTION:
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 24 SYMBOLS (CONT.) IPX1 Protection against ingress of water. NIBP Hose and Cuff C
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 25 SYMBOLS (CONT.) HISTORY VOLUME ALARM LIMITS ARROW UP ARROW D
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 26 SYMBOLS (CONT.) These symbols appear on the battery pack in place of text. Recyc
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 27 4. SAFETY MEASURES AND WARNINGS WARNING: The CARDELL MODEL 9401 and 9402 monitors are
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 28 WARNING: The use of accessory equipment not complying with the equivalent safety requireme
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 29 WARNING: ACCURACY – If the accuracy of any measurement does not seem reasonable, first che
21-02-0285 REV. 06 07/11 CARDELL® Veterinary Vital Signs Monitors MODEL DESCRIPTION 9401 Non-invasive Blood Pressure and Pulse Rate. 9402
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 30 CAUTION: Before each use, make sure that the monitor default alarm settings are appropriat
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 31 CAUTION: ACCIDENTAL SPILLS – In the event that fluids are accidentally spilled on the moni
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 32 GENERAL NOTES: There are no known risks with common disposal of equipment or accessories;
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 33 5. BLOOD PRESSURE MONITORING CUFF SELECTION AND APPLICATION The use of properly desig
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 34 SITE SELECTION Place the patient on a padded surface or chair to provide comfort. Shiv
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 35 CUFF PLACEMENT FOR A DOG For measurements in dogs, it is preferable to use the right lat
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 36 LARGER ANIMALS A large animal such as a horse should be in a stock, standing still, or ly
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 37 Use the following table as a guide to select the correct size. Table 2: Small Animal Cuff
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 38 Normal feline blood pressure: 124/84 Other investigators have reported comparable refere
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 39 GUIDELINES2 Mean Arterial Pressure (MAP): Minimum to adequately perfuse all peripheral ti
21-02-0285 REV. 06 07/11 In the U.S. the following caution applies: CAUTION: Federal law restricts this device to sale by or on the order of a v
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 40 This page is intentionally left blank NextBackGo To Table Of Contents
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 41 6. PULSE OXIMETRY MONITORING (Model 9402 only) The CARDELL Model 9402 Monitor is intend
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 42 7) Be sure that the sensor cable is positioned along the side of the animal’s face and bo
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 43 SENSOR REMOVAL CAUTION: For the comfort of the patient and to avoid damaging the sensor,
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 44 PERFORMANCE CONSIDERATIONS If there is excessive ambient light, cover the sensor site wit
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 45 7. MONITOR OPERATION FRONT PANEL Figure 6: Front Panel View DIGITAL DISPLAY AND
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 46 ♥ BPM Red colored LEDs indicate the Pulse Rate in BPM (beats per minute). Green colore
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 47 FRONT PANEL CONTROLS Figure 7: Front Panel Controls ON/STANDBY: Press once, turns
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 48 START/STAT: START: Initiates a blood pressure measurement in the Manual Mode or begins
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 49 VOLUME: Used to set the volume level of the Alarms and the SpO2 Beep (Model 9402). Use
21-02-0285 REV. 06 07/11 Manufacturers Declaration of Conformity Electronic Emissions and Immunity The Model 9401/9402 Monitor is intended for us
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 50 REAR PANEL Figure 8: Rear Panel View Monitor shown with optional RS232 Communications
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 51 LEFT SIDE VIEW Figure 9: Left Side Panel View CUFF HOSE CONNECTION The six (6)
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 52 MONITOR OPERATING INSTRUCTIONS TURNING THE CARDELL MONITOR “ON” Press the ON/STANDBY p
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 53 NOTE: If the printer was powered “ON” during the monitor’s power-up sequence, the Midmark
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 54 MANUAL MODE FOR BLOOD PRESSURE DETERMINATION Select the appropriate sized cuff for the
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 55 Press the CANCEL pushbutton, at any time, to stop a measurement and deflate the cuff durin
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 56 STAT MODE WARNING: Readings obtained during STAT mode may not meet the stated accuracy
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 57 HISTORY MODES The History Mode allows the user to recall previously taken measurements.
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 58 EVENT HISTORY (EVENT-Ev) To review History based on Event readings, press the HISTORY pus
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 59 PRINT HISTORY (to optional printer) The stored Event History and Trend History data are
21-02-0285 REV. 06 07/11 Guidance and Manufacturer’s Declaration – Electromagnetic Immunity The Model 9401/9402 Monitor is intended for use in the
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 60 PATIENT ALARM MODE WARNING: Configuring the CARDELL Monitor’s alarm settings to “OFF” w
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 61 The Message Window will display “Saving” and the current value(s) will be saved. When fin
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 62 AUDIBLE AND VISUAL INDICATORS The CARDELL Monitor is capable of producing both an audibl
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 63 CLEARING ALARMS The CARDELL Monitor provides to the user with an audible and visual indic
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 64 EQUIPMENT ALARMS During an Equipment Alarm, the monitor displays the alarm parameter in e
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 65 2-MINUTE AUDIO SILENCE NOTE: Enabling the 2-Minute Audio Alarm Off mode can only be acc
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 66 When enabled, the monitor will not alarm for patient related High and Low alarms, NIBP App
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 67 MONITOR CONFIGURATION The Monitor Configuration section allows the user to configure the
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 68 SOFTWARE REVISIONS The CARDELL Monitor displays the current software revision of its oper
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 69 To configure the alarms, first enter the Monitor Configuration menu. Once in the menu, us
21-02-0285 REV. 06 07/11 WARRANTY POLICY MONITORS CAS Medical Systems, Inc. warrants the monitor, when new, to be free from defects in material
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 70 MAP VALUE ENABLE / DISABLE During a blood pressure reading, the user can elect to display
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 71 SETTING THE TIME The CARDELL Monitor’s Time value is set for Eastern Time and is set at
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 72 BATTERY POWER The CARDELL Monitor is equipped with an internal rechargeable battery. T
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 73 CHECKING BATTERY STATUS NOTE: The monitor, when connected to AC Line Power and is powered
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 74 For the Auto “OFF” feature to work properly, the following conditions must apply; Alrm
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 75 SpO2 USER MESSAGES (Model 9402 only) If the accuracy of any measurement does not seem rea
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 76 ERROR MESSAGES ON THE MESSAGE WINDOW Table 7: Error Messages on the Message Window ERR
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 77 ERROR MESSAGES ON THE MESSAGE WINDOW (cont.) Table 7: Error Messages on the Message Win
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 78 ERROR MESSAGES ON THE MESSAGE WINDOW (cont.) Table 7: Error Messages on the Message Win
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 79 8. EXTERNAL PRINTER (optional) The following section is provided as an overview of the C
21-02-0285 REV. 06 07/11 This page is intentionally left blank NextBackGo To Table Of Contents
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 80 PRINTER CONTROLS AND INDICATORS Figure 11: Printer Controls and Indicators LED Ind
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 81 PRINTER OPERATION DIRECT CONNECTION The monitor uses the optional DB9 connector, located
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 82 Sample printouts of both History Modes are shown below. Figure 12: Sample Printouts 940V
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 83 CHARGING THE PRINTER BATTERY The printer is equipped with a rechargeable Lithium Ion (LiI
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 84 INSTALLING PAPER NOTE: A red line appears when the remaining supply of thermal paper beco
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 85 REPLACING THE BATTERY PACK WARNING: Do not operate the printer or connect the printer to
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 86 INSTALLING A NEW BATTERY PACK 1) Connect the battery cable into the battery connector
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 87 9. CLEANING CLEANING OVERVIEW WARNING: Do not, under any circumstances, perform any tes
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 88 THE DISPLAY CAUTION: Use care when cleaning the display. Scratches may occur. Occasio
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 89 PRINTER When the printer becomes dirty, wipe with a soft dry cloth. For extreme dirt b
21-02-0285 REV. 06 07/11 HOW TO CONTACT US For Warranty Issues: For Product Usage Information: CAS Medical Systems, Inc. Midmark 44 East Indu
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 90 This page is intentionally left blank NextBackGo To Table Of Contents
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 91 10. MAINTENANCE MAINTENANCE INTERVALS Besides the routine cleaning of the monitor and a
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 92 TEST MODE WARNING: Do not place the monitor in the TEST MODE when a cuff is attached to
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 93 LED CHECK The CARDELL Monitor incorporates an all lights “ON” check to verify the functio
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 94 6) Use the manometer inflation bulb to slowly inflate the system pausing for 30 seconds a
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 95 5) The Message Window will display “Chk Pres”, will inflate to approximately 180 mmHg and
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 96 OXIMETRY CALIBRATION CHECK The oximeter is factory calibrated to determine the percenta
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 97 INSTALLING THE BATTERY 1) Align the Battery Pack guides with the bottom of the monitor.
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 98 To replace the fuses: 1) Turn the monitor “OFF” and disconnect the power cord from the b
CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 99 11. ACCESSORIES BLOOD PRESSURE CUFFS Reusable Blood Pressure Cuffs (single tube) Ca