Midmark CARDELL 9402 User Manual

Browse online or download User Manual for Blood pressure units Midmark CARDELL 9402. Midmark CARDELL 9402 User`s manual

  • Download
  • Add to my manuals
  • Print

Summary of Contents

Page 1 - User’s Manual

21-02-0285 REV. 06 07/11 CARDELL® Veterinary Vital Signs Monitors Models 9401 and 9402 User’s Manual Manufactured for: Go To Table

Page 2

21-02-0285 REV. 06 07/11 This page is intentionally left blank NextBackGo To Table Of Contents

Page 3

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 100 OTHER ACCESSORIES: Catalog No. Description 01-02-0395 Replacement Power Cord, U.S.A.

Page 4 - CAUTION:

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 101 MONITOR CONFIGURATIONS Table 8: Monitor Configurations Model Description 9401 MAXNIBP

Page 5

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 102 This page is intentionally left blank NextBackGo To Table Of Content

Page 6

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 103 12. GLOSSARY Diastolic The Diastolic pressure is the bottom (lower) number given in a

Page 7

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 104 This page is intentionally left blank NextBackGo To Table Of Contents

Page 8

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 105 13. SPECIFICATIONS NIBP MEASUREMENT Characteristic Specification Technique:Oscillom

Page 9

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 106 PATIENT ALARMS 9401 9402 Patient Parameters Low Limit High Limit Low Limit Hi

Page 10 - Go To Table Of Contents

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 107 POWER Characteristic Specification Source: External line or internal battery AC Power

Page 11 - TABLE OF CONTENTS

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 108 OPERATING ENVIRONMENT Characteristic Specification Operating Temperature: 0°C to 50°C (

Page 12

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 109 STANDARDS Units comply with the following requirements: IEC 60601-1 EN 60601-1-2 EN 6060

Page 13

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 11 TABLE OF CONTENTS 1. INTRODUCTION AND INTENDED USE ...

Page 14

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 110 This page is intentionally left blank NextBackGo To Table Of Contents

Page 15 - FIGURES

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 111 14. PURCHASING RECORD CARDELL VETERINARY MONITOR Model Purchased (select one): 9401

Page 16

B-26© Midmark Corporation 2009Midmark Corporation60 Vista DriveVersailles, OH 45380-0286Phone: 937-526-3662Fax: 937-526-5542midmark.comBackGo To Tab

Page 17

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 12 REAR PANEL ...

Page 18

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 13 MONITOR CONFIGURATION ...

Page 19

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 14 PNEUMATIC TUBING ...

Page 20

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 15 FIGURES Figure 1: Patient Environment ...

Page 21 - 2. UNPACKING THE MONITOR

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 16 TABLES Table 1: Parts of the System ...

Page 22

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 17 1. INTRODUCTION AND INTENDED USE INTRODUCTION The CARDELL Veterinary Monitors, Models 9

Page 23 - 3. SYMBOLS

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 18 On each detected pulse, the perfusion LED does indicate patient perfusion signals. This b

Page 24

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 19 DEFINITION OF TERMS In this manual, “WARNING”, “CAUTION”, “IMPORTANT” and “NOTE” mean the

Page 25 - ♥ BPM Pulse Rate Display

21-02-0285 REV. 06 07/11 NextBackGo To Table Of Contents

Page 26

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 20 This page is intentionally left blank NextBackGo To Table Of Contents

Page 27

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 21 2. UNPACKING THE MONITOR INITIAL INSPECTION Before unpacking the monitor, inspect the

Page 28

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 22 This page is intentionally left blank NextBackGo To Table Of Contents

Page 29

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 23 3. SYMBOLS Units may display the following symbols: Alternating Current CAUTION:

Page 30

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 24 SYMBOLS (CONT.) IPX1 Protection against ingress of water. NIBP Hose and Cuff C

Page 31

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 25 SYMBOLS (CONT.) HISTORY VOLUME ALARM LIMITS ARROW UP ARROW D

Page 32

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 26 SYMBOLS (CONT.) These symbols appear on the battery pack in place of text. Recyc

Page 33

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 27 4. SAFETY MEASURES AND WARNINGS WARNING: The CARDELL MODEL 9401 and 9402 monitors are

Page 34

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 28 WARNING: The use of accessory equipment not complying with the equivalent safety requireme

Page 35

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 29 WARNING: ACCURACY – If the accuracy of any measurement does not seem reasonable, first che

Page 36

21-02-0285 REV. 06 07/11 CARDELL® Veterinary Vital Signs Monitors MODEL DESCRIPTION 9401 Non-invasive Blood Pressure and Pulse Rate. 9402

Page 37

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 30 CAUTION: Before each use, make sure that the monitor default alarm settings are appropriat

Page 38

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 31 CAUTION: ACCIDENTAL SPILLS – In the event that fluids are accidentally spilled on the moni

Page 39

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 32 GENERAL NOTES: There are no known risks with common disposal of equipment or accessories;

Page 40

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 33 5. BLOOD PRESSURE MONITORING CUFF SELECTION AND APPLICATION The use of properly desig

Page 41

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 34 SITE SELECTION Place the patient on a padded surface or chair to provide comfort. Shiv

Page 42

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 35 CUFF PLACEMENT FOR A DOG For measurements in dogs, it is preferable to use the right lat

Page 43

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 36 LARGER ANIMALS A large animal such as a horse should be in a stock, standing still, or ly

Page 44

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 37 Use the following table as a guide to select the correct size. Table 2: Small Animal Cuff

Page 45 - 7. MONITOR OPERATION

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 38 Normal feline blood pressure: 124/84 Other investigators have reported comparable refere

Page 46

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 39 GUIDELINES2 Mean Arterial Pressure (MAP): Minimum to adequately perfuse all peripheral ti

Page 47

21-02-0285 REV. 06 07/11 In the U.S. the following caution applies: CAUTION: Federal law restricts this device to sale by or on the order of a v

Page 48

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 40 This page is intentionally left blank NextBackGo To Table Of Contents

Page 49

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 41 6. PULSE OXIMETRY MONITORING (Model 9402 only) The CARDELL Model 9402 Monitor is intend

Page 50

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 42 7) Be sure that the sensor cable is positioned along the side of the animal’s face and bo

Page 51

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 43 SENSOR REMOVAL CAUTION: For the comfort of the patient and to avoid damaging the sensor,

Page 52

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 44 PERFORMANCE CONSIDERATIONS If there is excessive ambient light, cover the sensor site wit

Page 53

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 45 7. MONITOR OPERATION FRONT PANEL Figure 6: Front Panel View DIGITAL DISPLAY AND

Page 54

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 46 ♥ BPM Red colored LEDs indicate the Pulse Rate in BPM (beats per minute). Green colore

Page 55

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 47 FRONT PANEL CONTROLS Figure 7: Front Panel Controls ON/STANDBY: Press once, turns

Page 56

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 48 START/STAT: START: Initiates a blood pressure measurement in the Manual Mode or begins

Page 57

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 49 VOLUME: Used to set the volume level of the Alarms and the SpO2 Beep (Model 9402). Use

Page 58

21-02-0285 REV. 06 07/11 Manufacturers Declaration of Conformity Electronic Emissions and Immunity The Model 9401/9402 Monitor is intended for us

Page 59

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 50 REAR PANEL Figure 8: Rear Panel View Monitor shown with optional RS232 Communications

Page 60

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 51 LEFT SIDE VIEW Figure 9: Left Side Panel View CUFF HOSE CONNECTION The six (6)

Page 61

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 52 MONITOR OPERATING INSTRUCTIONS TURNING THE CARDELL MONITOR “ON” Press the ON/STANDBY p

Page 62

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 53 NOTE: If the printer was powered “ON” during the monitor’s power-up sequence, the Midmark

Page 63

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 54 MANUAL MODE FOR BLOOD PRESSURE DETERMINATION Select the appropriate sized cuff for the

Page 64

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 55 Press the CANCEL pushbutton, at any time, to stop a measurement and deflate the cuff durin

Page 65

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 56 STAT MODE WARNING: Readings obtained during STAT mode may not meet the stated accuracy

Page 66

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 57 HISTORY MODES The History Mode allows the user to recall previously taken measurements.

Page 67

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 58 EVENT HISTORY (EVENT-Ev) To review History based on Event readings, press the HISTORY pus

Page 68

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 59 PRINT HISTORY (to optional printer) The stored Event History and Trend History data are

Page 69

21-02-0285 REV. 06 07/11 Guidance and Manufacturer’s Declaration – Electromagnetic Immunity The Model 9401/9402 Monitor is intended for use in the

Page 70

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 60 PATIENT ALARM MODE WARNING: Configuring the CARDELL Monitor’s alarm settings to “OFF” w

Page 71

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 61 The Message Window will display “Saving” and the current value(s) will be saved. When fin

Page 72

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 62 AUDIBLE AND VISUAL INDICATORS The CARDELL Monitor is capable of producing both an audibl

Page 73

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 63 CLEARING ALARMS The CARDELL Monitor provides to the user with an audible and visual indic

Page 74

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 64 EQUIPMENT ALARMS During an Equipment Alarm, the monitor displays the alarm parameter in e

Page 75

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 65 2-MINUTE AUDIO SILENCE NOTE: Enabling the 2-Minute Audio Alarm Off mode can only be acc

Page 76

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 66 When enabled, the monitor will not alarm for patient related High and Low alarms, NIBP App

Page 77

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 67 MONITOR CONFIGURATION The Monitor Configuration section allows the user to configure the

Page 78

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 68 SOFTWARE REVISIONS The CARDELL Monitor displays the current software revision of its oper

Page 79

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 69 To configure the alarms, first enter the Monitor Configuration menu. Once in the menu, us

Page 80

21-02-0285 REV. 06 07/11 WARRANTY POLICY MONITORS CAS Medical Systems, Inc. warrants the monitor, when new, to be free from defects in material

Page 81

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 70 MAP VALUE ENABLE / DISABLE During a blood pressure reading, the user can elect to display

Page 82

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 71 SETTING THE TIME The CARDELL Monitor’s Time value is set for Eastern Time and is set at

Page 83

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 72 BATTERY POWER The CARDELL Monitor is equipped with an internal rechargeable battery. T

Page 84

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 73 CHECKING BATTERY STATUS NOTE: The monitor, when connected to AC Line Power and is powered

Page 85

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 74 For the Auto “OFF” feature to work properly, the following conditions must apply;  Alrm

Page 86

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 75 SpO2 USER MESSAGES (Model 9402 only) If the accuracy of any measurement does not seem rea

Page 87 - 9. CLEANING

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 76 ERROR MESSAGES ON THE MESSAGE WINDOW Table 7: Error Messages on the Message Window ERR

Page 88

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 77 ERROR MESSAGES ON THE MESSAGE WINDOW (cont.) Table 7: Error Messages on the Message Win

Page 89

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 78 ERROR MESSAGES ON THE MESSAGE WINDOW (cont.) Table 7: Error Messages on the Message Win

Page 90

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 79 8. EXTERNAL PRINTER (optional) The following section is provided as an overview of the C

Page 91 - 10. MAINTENANCE

21-02-0285 REV. 06 07/11 This page is intentionally left blank NextBackGo To Table Of Contents

Page 92

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 80 PRINTER CONTROLS AND INDICATORS Figure 11: Printer Controls and Indicators  LED Ind

Page 93

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 81 PRINTER OPERATION DIRECT CONNECTION The monitor uses the optional DB9 connector, located

Page 94

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 82 Sample printouts of both History Modes are shown below. Figure 12: Sample Printouts 940V

Page 95

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 83 CHARGING THE PRINTER BATTERY The printer is equipped with a rechargeable Lithium Ion (LiI

Page 96

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 84 INSTALLING PAPER NOTE: A red line appears when the remaining supply of thermal paper beco

Page 97

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 85 REPLACING THE BATTERY PACK WARNING: Do not operate the printer or connect the printer to

Page 98

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 86 INSTALLING A NEW BATTERY PACK 1) Connect the battery cable into the battery connector

Page 99 - 11. ACCESSORIES

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 87 9. CLEANING CLEANING OVERVIEW WARNING: Do not, under any circumstances, perform any tes

Page 100

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 88 THE DISPLAY CAUTION: Use care when cleaning the display. Scratches may occur. Occasio

Page 101

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 89 PRINTER When the printer becomes dirty, wipe with a soft dry cloth. For extreme dirt b

Page 102

21-02-0285 REV. 06 07/11 HOW TO CONTACT US For Warranty Issues: For Product Usage Information: CAS Medical Systems, Inc. Midmark 44 East Indu

Page 103 - 12. GLOSSARY

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 90 This page is intentionally left blank NextBackGo To Table Of Contents

Page 104

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 91 10. MAINTENANCE MAINTENANCE INTERVALS Besides the routine cleaning of the monitor and a

Page 105 - 13. SPECIFICATIONS

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 92 TEST MODE WARNING: Do not place the monitor in the TEST MODE when a cuff is attached to

Page 106

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 93 LED CHECK The CARDELL Monitor incorporates an all lights “ON” check to verify the functio

Page 107

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 94 6) Use the manometer inflation bulb to slowly inflate the system pausing for 30 seconds a

Page 108

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 95 5) The Message Window will display “Chk Pres”, will inflate to approximately 180 mmHg and

Page 109

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 96 OXIMETRY CALIBRATION CHECK The oximeter is factory calibrated to determine the percenta

Page 110

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 97 INSTALLING THE BATTERY 1) Align the Battery Pack guides with the bottom of the monitor.

Page 111 - 14. PURCHASING RECORD

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 98 To replace the fuses: 1) Turn the monitor “OFF” and disconnect the power cord from the b

Page 112

CARDELL® Veterinary Monitors 21-02-0285 REV. 06 07/11 99 11. ACCESSORIES BLOOD PRESSURE CUFFS Reusable Blood Pressure Cuffs (single tube) Ca